UIP> za 45
tpick% DSOSinit
p
. tpick% DSOSpower on
1
.tpick% DSOSagent
agent.log
.ctrl-C
.tpick% DSOSbase
tpick% DSOSpower off
During observations, please check the system periodically.
Should any of these faults occur, turn off the power supplies first. Then shut down the rest of the system.
Do not use the DSOS until the fault is repaired.
Left uncorrected, these faults may cause system damage and misadjustment of the dish panels.
Essential. Please notify CSO staff what failed and when.
If the system was initialized during the day before sunset, reinitializing 1 hour after the sunset is worthwhile. Another reason for reinitialize is a change of more than 5 °C in the ambient temperature since the original initialization. A significant temperature change in the middle of the night often indicates, however, bad observing weather, so be sure to check for rain or snow. Shutting the dome may be necessary. It is not recommended to rebaseline often.
ctrl-C
in the agent window. DSOSbase
in the command window. DSOSpower off
in the command window. UIP> za 45
).On occasion, the power supplies get confused, which prevents them from turning on correctly. If this occurs, reset them.
p
. DSOSpower off
in the command window.DSOSpower reset
in the command window. OFFLINE
and the times won't update.DSOSpower off
in the command window.STANDBY
and the times update again.DSOSpower on
in the command window. Do not repeat this procedure more than a couple of times. If it doesn't fix the problem, then something else must be wrong.
In unusual circumstances, power supplies may require manual starts. Power supplies are always operated in pairs.
Output 0.00 VOLTS
0.0 AMPS
MENU
buttons near the keypad on the right of the front panels. Restore from Memory: N
15.00 V 75.0 A 3000 W
ENTER
buttons on the right of the panels. This applies both biases to the power amps at the same time.Output 15.00 VOLTS
75.0 AMPS
ENABLE
on the left side. ENTER
if a mixed configuration is used.If any of the chillers sounds an alarm, check its error display.
ADD
: Error 54
: If any of the four pumps supplying the dish manifolds fails, usually the meters will read low (or no) pressure and the pump may be hot to the touch.